*This bulletin supersedes TB 9-6625-2021-35, dated 20 June 1997.
*TB 9-6625-2021-35
SUPERSEDED COPY DATED 20 JUNE 1997
DEPARTMENT OF THE ARMY TECHNICAL BULLETIN
CALIBRATION PROCEDURE FOR
FREQUENCY COUNTER
HEWLETT-PACKARD, MODEL 5340A
AND
MICROWAVE FREQUENCY COUNTER
TD-1225A(V)1/U (HEWLETT-PACKARD, MODEL 5342A/H14),
TD-1225A(V)2/U (HEWLETT-PACKARD, MODEL 5342A/H16) AND
HEWLETT-PACKARD, MODELS 5342A AND 5343A
Headquarters, Department of the Army, Washington, DC
26 June 2000
Approved for public release, distribution is unlimited.
REPORTING OF ERRORS
You can help improve this publication by calling attention to errors and by
recommending improvements and stating your reasons for the
recommendations. Your letter or DA Form 2028, Recommended Changes to
Publications, should be mailed directly to Commander, U. S. Army Aviation and
Missile Command, ATTN: AMSAM-TMD-LP, Redstone Arsenal, AL 35898-5000.
FAX to DSN 788-2313 (Commercial: 256-842-2313),. A reply will be furnished
directly to you.
Paragraph
Page
SECTION I.
IDENTIFICATION AND DESCRIPTION
Test instrument identification ..........................
1 2
Forms, records, and reports .............................
2 3
Calibration description .....................................
3 3
II.
EQUIPMENT REQUIREMENTS
Equipment required .........................................
4 4
Accessories required .........................................
5 4
III.
CALIBRATION PROCESS FOR FREQUENCY
COUNTER HEWLETT-PACKARD, MODEL
5340A
Preliminary instructions ................................
6 5