TB 9-6625-2021-35 
13 
 (4) Allow at least 24 hours for TI time base oscillator to stabilize.  Frequency 
difference meter indication  will be within  ±5 parts in 10
10 
of indication recorded in (3) 
above.  Record frequency difference meter indication. 
 (5) Adjust autotransformer output to 105 V ac.  Allow 1 minute for oscillator to 
stabilize.  Frequency difference meter indication will be within ±1 part in 10
10
 of indication  
recorded in (4) above. 
 (6) Adjust autotransformer output to 125 V ac.  Allow 1 minute for oscillator to 
stabilize.  Frequency difference meter indication will be within ±1 part in 10
10  
of indication 
action  recorded in (4) above. 
 (7) Adjust autotransformer output to 115 V ac. 
 b. Adjustments. No further adjustments can be made. 
17. Self-Check 
 a. Performance Check 
 (1) Press FREQ SHIFT (blue) and RESOLUTION CHECK pushbuttons.  Verify 
that TI displays 75.000000 MHz ±1 count.  
 (2) Press RESET pushbutton. 
 (3) Remove 50Ω termination from FREQ STD OUT (rear panel) and connect 
FREQ STD OUT to TI front panel BNC input connector. 
 (4) Press FREQ SHIFT (blue) and RESOLUTION 1 Hz pushbuttons.  Verify that 
TI displays 10.000000 MHz ±1 count. 
 (5) Press RESET pushbutton and reconnect 50Ω termination to FREQ STD OUT 
connector (rear panel). 
 b. Adjustments.  No adjustments can be made. 
18. Frequency and Sensitivity Check - 10 Hz to 520 MHz  
 a. Performance Check 
CAUTION 
The 10 Hz to 500 MHz direct count input BNC connector is 
fuse-protected for a maximum input level of 3.5 V rms (+24 
dBm)  
 (1) Press FREQ SHIFT (blue) and RESOLUTION 1 Hz pushbutton. 
 (2) Connect function generator Function Outputs Unbalanced to TI front panel 
BNC input. 
 (3) Set function generator for a sinewave, 10 Hz, 1 mVp-p, 50Ω output.  Slowly 
increase function generator amplitude until TI displays a stable indication of applied frequency.  
If function generator amplitude exceeds 71 mVp-p, perform b below.