Increasing Measurement Throughput
Optimizing Measurement Setup Time
Measurement setup time is defined as the time required to configure an individual
instrument within the Test Set to make a measurement.
In general there are two methodologies which can be used to setup individual
instruments in the Test Set:
1. Set up every field every time a measurement is made.,
2. Define a base instrument state and then modify it as needed for each measurement
(always returning to the base state after finishing the measurement).
Defining a base instrument state requires fewer HP-IB transactions to set up an
instrument (in the majority of cases) which in turn reduces measurement setup
time.