EasyManua.ls Logo

HP 8970B - General Adjustment Overview; Input Power Detector Calibration

HP 8970B
40 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3-2 HP 8970B Option 020 Service Manual Supplement
Adjustments
General
General
There are 2 adjustments that need to be made:
1. Input power detector gain and offset.
2. The 600 MHz SAW oscillator frequency adjustment.
Input power detector Gain and Offset Adjustment
Follow the adjustment procedure in section 5-13 of the Adjustments Chapter in the 8970B
Service Manual with the following exceptions.
Figure 3-1: Input detector gain and offset adjustment
1. The input level must be -21 dBm ± 0.1 dB.
2. R32 is the offset adjustment.
3. R26 is the gain adjustment.
R32
R26
J1
RF Input
Assembly