HP ESG Family Signal Generators Getting Started
Creating an FM Signal
Creating an FM Signal
This procedure will show you how to configure the signal generator to output a
frequency-modulated signal with the following characteristics:
l
carrier frequency set to 104.9 MHz
l
power level set to 0 dBm
l
FM deviation set to 75 kHz
l
FM rate set to 10 kHz
Setting the Carrier Frequency
1.
Press Preset. This sets the signal generator to its factory-defined instrument
state.
2. Press Frequency.
Frequency becomes the active function and the preset value for
frequency is displayed in the active entry area.
3. Enter 104.9 using the numeric The new carrier frequency (10
4 .9 0 0 0 0 0 0 0
MHZ) is
keypad and press the MHz
shown in both the FREQUENCY area and the active entry area
softkey.
of the display.
Setting the Power Level
1.
Press Amplitude.
Amplitude becomes the active function and the preset value for
amplitude is displayed in the active entry area.
2. Enter
0
using the numeric
The new power level (0 . 0 0 dBm) is shown in both the
keypad and press the dBm
AMPLITUDE area and the active entry area of the display.
softkey.
Quick Start Guide