CAUTION: Replace fuse(s) as marked. All
fuses must be replaced with a fuse of the same
size and rating. Refer to the Wire Diagrams
at the end of this Manual.
PRÉCAUTION : Remplacez le(s) fusible(s)
comme indiqué. Tous les fusibles doivent
être remplacés par un fusible de la même
dimension et de la même valeur. Se référer
aux schémas de câblage à la fin de ce manuel
The chair contains two fuses located inside the AC Input
Module in the Base Assembly.
4.1. Fuse Replacement (AC Input Module)
4.1.1. Unplug the power cord from the AC power input
module. Referring to Figure 7, locate tab. With a tool similar
to a small common screwdriver placed under tab, pry fuse
drawer out. Remove the drawer by pulling it out with your
fingers.
4.1.2. To remove fuses from fuse drawer, tilt fuses in direction
of arrows as shown in Figure 7. Examine the fuses, replace
as necessary.
FIGURE 7
AC INPUT MODULE
CAUTION: To avoid a bind of the Fuse Drawer
within the AC Input Module, never install the
Fuse Drawer into the AC Input Module without
fuses.
PRÉCAUTION : Pour éviter que le tiroir à
fusible reste bloqué dans le module d’entrée
CA, ninstallez jamais le tiroir à fusible
à l’intérieur du module dente CA sans
fusibles.
3.7.6. Adjustment-No. 11, 21 Headrest
3.7.6.1. The handle on the outermost knob can be adjusted
to any preferred, locked position. Simply push in on the large
screw while pulling out on the handle. Rotate as desired, and
release. (See Figure 6, No. 11 Headrest shown)
3.7.6.2. The No. 21 headrest pillow assembly will slide
to a preferred patient position. Simply slide pillow on the
attached headrest strap to the desired position. (Not Shown)
FIGURE 6
No. 11 Headrest
4.0 LOCATING AND REPLACING FUSES
WARNING: DISCONNECT EQUIPMENT FROM
MAIN INPUT POWER BEFORE PROCEEDING WITH
ELECTRICAL INSPECTIONS OR MAINTENANCE.
AVERTISSEMENT : DÉCONNECTEZ
LÉQUIPEMENT DE LALIMENTATION
ÉLECTRIQUE PRINCIPALE AVANT DE
PROCÉDER À DES INSPECTIONS ÉLECTRIQUES
OU À DE L’ENTRETIEN.
HANDLE
SCREW

IN-6200
11