Status bits for the Outer Dataset® Arm
Unused bits will be “0”. For example, if calibration passed, the bits will be “000111”.
The bits are described below.
Bit 1 - This bit checks the readings in lateral steps 1 and 2 of the outer
Dataset® calibration. If this bit fails, there is something wrong
with the outer Dataset® distance sensor circuit or the correct
calibration sequence was not followed.
Bit 2 - This bit checks the readings in lateral steps 2 and 3 of the outer
Dataset® calibration. If this bit fails, there is something wrong
with the outer Dataset® distance sensor circuit sensor circuit or
the correct calibration sequence was not followed.
Bit 3 - This bit checks the readings in lateral step 1 and the initial
storing of the home settings of the outer Dataset® calibration. If
this bit fails, there is something wrong with the outer Dataset®
distance sensor circuit or the correct calibration sequence was
not followed.
DSP9600/9100 Service Manual Calibration Results 15