Product Description
DOCUMENT HQI-J-FC1-001
ES_DATE 2019-07-19
Fundus camera
HFC-1
RE_DATE 2020-06-08
REV R-3 PAGE 24 / 36
Q-I-300-02(R1):2018/06/04 Huvitz Co.,Ltd
Dispersion
Compensation
Mode for checking dispersion compensation parameters
Spectrometer
Calibration
Mode for acquiring the parameters of K-Linearization checking raw signal data
with spectrometer calibrator attached.
Kerato
Calibration
Mode for acquiring Kerato focus parameters from the mire ring image on anterior
IR camera, and checking the quality and optimization of split focus signal on
retinal IR camera.
Color Camera
Test
Mode for checking the static image and live images from fundus camera.
System
Configuration
Mode for checking and storing the parameters for system initialization, such as
the serial number of the device, the location of the motor, the spectrometer
setting, and the duty of the auto tracking motor
Factory Setup
Mode for checking and storing the parameters for operation, such as the initial
value of system LED, IR camera settings, parameter of fundus image and the
fixation's position
Packing Mode Move Chinrest and the body of HOCT the lowest position
Exit Exit the Device Calibrator menu.
9.2 System Configuration