WARNING!
• Portable RF communications equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm (12 inches) to any part of the Normatec Elite, including cables
specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
• The Normatec Elite should not be used adjacent to or stacked with other equipment. If adjacent
or stacked use is necessary, the Normatec Elite should be observed to verify normal operation. If
operation is not normal, the Normatec Elite or the other equipment should be moved.
• Use of accessories, transducers and cables other than those specified or provided by the manufacturer
of this equipment could result in increased electromagnetic emissions or decreased electromagnetic
immunity of this equipment and result in improper operation.
• Avoid exposure to known sources of EMI (electromagnetic interference) such as diathermy, lithotripsy,
electrocautery, RFID (Radio Frequency Identification), and electromagnetic security systems such as anti-
theft/electronic article surveillance systems, metal detectors. Note that the presence of RFID devices may not
be obvious. If such interference is suspected, reposition the equipment, if possible, to maximize distances.
ELECTROMAGNETIC EMISSIONS
This equipment is intended for use in the electromagnetic environment of clinics, hospitals, athlete training,
or home environments. The user of this equipment should assure that it is used in such an environment.
Emissions
Compliance
According To
Electromagnetic Environment
RF emissions
(CISPR 11)
Group 1 The equipment uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
CISPR emissions
classification
Class B The equipment is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage charger network
that supplies buildings used for domestic purposes.
Harmonic emissions
(IEC 61000-3-2)
Class A
Voltage fluctuations/
flicker (IEC 61000-3-3)
Complies
ELECTROMAGNETIC IMMUNITY
During the immunity testing described below the Normatec Elite continued to provide therapy normally.
This equipment is intended for use in the electromagnetic environment specified below. The user of this
equipment should assure that it is used in such an environment.
Immunity
Against
Compliance
Level (of this
control unit)
Electromagnetic Environment
Electrostatic
discharge, ESD
(IEC 61000-4-2)
± 8 kV Direct
± 2,4,8,15 kV
Floors should be wood, concrete, or ceramic tile. If floors
are covered with synthetic material, the relative humidity
should be kept at levels to reduce electrostatic charge to
suitable levels.
Electrical fast
transients/bursts
(IEC 61000-4-4)
± 2 kV Mains power quality should be that of a typical clinic,
hospital, athletic training, or home environment.
RF Proximity
(IEC 61000-4-3)
27 V/m 28 V/m
9 V/m 28 V/m
28 V/m 28 V/m
9 V/m
Equipment with high RF emissions should be kept at a
distance to reduce the likelihood of interference.