CoBrite User Manual
10
2.5 CBDX2 Top side of unit
Note This unit utilizes the top cover for heat dissipation of internal components. Do not
reduce the thermal dissipation by blocking the airflow in this area by placing items
on it to avoid overheating of the unit.
2.6 CBDX2 Front side of unit
The front side is dominated by a large Touch Panel Color Display for intuitive control of the
instrument. The optical connector plate type depends on the laser configuration equipped on the
specific unit. The 2 screws allow to easily detach the front plate to allow removal and cleaning of
the connector.
Note The fibers attached to the plate are 250um buffered type and therefore sensitive to
breakage under mechanical stress. So, take care if removing the connector plate to
not damage optical fibers.
2.7 CBDX2 Rear side of unit
The rear side hosts electrical connectors that can be utilized by the user.
2.7.1 Power socket
Electrical power is supplied using a coaxial 12V DC jacket. Only use the supplied external
AC/DC adaptor to power the unit. The AC/DC adapter unit automatically adapts to local power
line specifications within the range and complies to standards printed on the adaptor.
2.8 CoBrite series - Common Ports
2.8.1 Trigger Ports
The SMA type sockets can be used to achieve timewise synchronization of the CoBrite with
other equipment. Both ports support 3.3V LV TTL standardized signals.
Trigger IN allows triggering tunable laser setting if the feature is enabled via Remote control or
GUI. Polarity and user set delays can be also configured this way. For positive polarity, the scan
is started with the positive flank und will be fully executed regardless of the trigger in status
afterwards. For negative polarity, the scan is triggered by the negative flank on the unit.