EasyManua.ls Logo

IDS Optex Xwave

IDS Optex Xwave
63 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Dip Switch
Dip Switch 1 - Walk Test Mode
Position
Function
ON
Test Mode On. The LED lights irrespective of the Dip switch 6 setting.
Dip switch 2, Battery Saving Timer, is inactive.
OFF
Normal Mode. LED and Battery Saving Timer works according to
their setting.
NB: Once the walk test is complete always set the detector to Normal Mode. Using the
detector in Test Mode will shorten the battery life.
Dip Switch 2 Battery Saving Timer
Position
Function
ON
Even if there are continuous alarms, the alarm is only generated once
in a 5 second time.
OFF
Even if there are continuous alarms, the alarm is only generated once
in a 120 second time. To lengthen battery life.
Dip Switch 3 Immunity
Position
Function
ON
Immunity logic is activated. Use this setting in harsher environments,
such as moving trees or bushes.
OFF
Normal Sensitivity.
Dip Switch 4 - Trouble Output
Position
Function
ON
Sets the output type to Normally Open (N.O.)
OFF
Sets the output type to Normally Closed (N.C.)
NB: This setting must be set to ON, or Anti-Masking will not be sent from the Wireless
Transmitter. Anti-Masking is sent as a tamper to the X-Series Panel.
Dip Switch 5 Low Battery Output
Position
Function
ON
Low Battery signal is sent to the Trouble Output.
OFF
Low battery output is not operational.
NB: The Wireless Transmitter monitors the battery and will send a Low Battery signal when the
battery gets low.
Dip Switch 6 LED
Position
Function
ON
Enables the LED.
OFF
Disables the LED. To lengthen battery life.