SECTION
5
Table
5-1.
Troubleshooting
Guide
(Continued)
Symptom/Step
Possible
Cause
Remarks
b.
Base
line
of
triangle
and
sine
offset
from
zero.
Misadjustment
of R228.
c.
No
dynamic
triggering.
Same
as
9a.
d.
No
manual
triggering.
Same
as
9b.
e.
Baseline
oscillations.
Start/Stop
trouble.
See
step
23.
11.
BURST MODE
FAULTS.
a.
No
burst
output.
1.
Open
circuit
from
TP27
to
R277
when
in
burst
mode.
2.
Same
as
9a.
b.
Continuous
output
only.
Same
as
9a.
12.
SWEEP
FAULTS
a.
Waveform
nonlinearity.
1.
Q61. Q64. Q65
leaky.
2.
CR73. CR74. CR75
leaky.
3.
C121-C125
leaky.
b.
Vernier
<100:1
sweep
time
range.
IC15
faulty.
c.
No
Sweep
Same
as
19.
13.
TRIGGERED
SWEEP
FAULTS
a.
No
sweep.
Same
as
19.
b.
Double
sweep.
Excessive
rate
on
TRIG/SYNC
input.
c.
Interrupted
sweep.
1.
Q63
failure.
2.
Logic
Failure.
See
step
22.
14.
OUTPUT
AMP
FAULTS
a.
Positive
Output
«
11
v
open
circuit)
at
all
times.
1.
Q33. Q17. CR31
or
CR32
open.
2.
Negative
supply
not
operating.
3.
VR7
shorted.
4.
IC9
faulty.
5-7