H
Heather RodgersAug 10, 2025
What to do if iHealth BG5 Blood Glucose Meter test results are inconsistent?
- MMeghan SchroederAug 10, 2025
If you are getting inconsistent test results from your iHealth Blood Glucose Meter, consider these potential causes and solutions: * **Insufficient Sample:** Ensure enough sample is applied to a new test strip when retesting. * **Expired Materials:** Retest using a new test strip or control solution if the current one has expired. * **Damaged Test Strip:** Perform a control solution test with a new test strip. If results remain out of range, switch to a new vial of test strips. This may occur if the test strip has been damaged due to heat or humidity, affecting sample application. * **Environmental Factors:** Allow the system to adjust to a room-temperature environment for about 30 minutes before conducting a new test.