Option 06 - 3mA AC/DC Current Limitation (all models)
This option limits the maximum current capability of the tester to 3mA AC or DC. The following specications table is applicable
when this option is installed in the tester:
AC WITHSTAND VOLTAGE
Settings
Hi-Limit AC current, mA Range:
Resolution:
Accuracy:
0.10 - 3.00
0.01
± (2% of setting + 2 counts)
Lo-Limit AC current, mA Range:
Resolution:
Accuracy:
0.00 - 3.00
0.01
± (2% of setting + 2 counts)
Measurement
AC Current, mA Range:
Resolution:
Accuracy:
0.10 - 3.00
0.01
± (2% of reading + 2 counts)
Short Circuit Protection minimum current 3.5 mA RMS at short circuit, Response time < 2ms
DC WITHSTAND VOLTAGE
Settings
Hi-Limit AC current, mA Range:
Resolution:
Accuracy:
0.00 - 3.00
0.01
± (2% of setting + 2 counts)
Lo-Limit AC current, mA Range:
Resolution:
Accuracy:
0.00 - 3.00
0.01
± (2% of setting + 2 counts)
Measurement
AC Current, mA Range:
Resolution:
Accuracy:
0.02 - 3.00
0.01
± (2% of reading + 2 counts)
Short Circuit Protection Minimum current 3.5 mA AVG at short circuit, Response time < 2ms