hipot.com 51
Test Parameter Editing Commands
These commands are used to modify the test parameter within each memory. These commands require a parameter value
to be included with the command. The companion query command will read the parameter. The writing of the parameter
requires that the unit not be included with the value, only the numeric value should be included with the command. Also when
the query commands are used the response will not include the units characters. Many of the commands will function the same
way for multiple test types however the input range may be dierent and therefore used a dierent possible set of values.
Command Name Test Types Values
EV <value>
EV?
Edit Voltage ACW
DCW
IR
0.00 – 5.00 (kV)
0.00 – 6.00 (kV)
100 – 1000 (V)
EH < value >
EH?
Edit HI-Limit ACW
DCW
IR
0.10 – 12.00 (mA)
0.02 – 5.00 (mA)
0, 1- 1000 (MΩ)
0 = O
EL < value >
EL?
Edit LO-Limit ACW
DCW
IR
0.10 – 12.00 (mA)
0.02 – 5.00 (mA)
0, 1- 1000 (MΩ)
0 = O
ERU < value >
ERU ?
Edit Ramp ACW
DCW
IR
0.2 – 180.0 (sec)
0.2 – 180.0 (sec)
.1, 2.0 (sec)
EDW < value >
EDW?
Edit Dwell ACW
DCW
0.0, 0.2 – 60.0s
0.0, 0.2 – 60.0s
0.0 = Continuous
EDE < value >
EDE?
Edit Delay IR 0.0, 0.5 - 999.9s
0.0 = Continuous
ECT {1|0}
ECT?
Edit Continuity ACW
DCW
1=On, 0=O
ECH < value >
ECH?
Edit Continuity HI-Limit ACW
DCW
0.00 – 1.50 (Ω)
ECL < value >
ECL?
Edit Continuity LO-Limit ACW
DCW
0.00 – 1.50 (Ω)
ECO < value >
ECO?
Edit Continuity Oset ACW
DCW
0.00 – 0.50 (Ω)
EF {1|0}
EF?
Edit Frequency ACW 1=60 Hz, 0=50 Hz
ECC {1|0}
ECC?
Edit Connect ACW
DCW
IR
1=On, 0=O