24
6.8.3 Termination of Charging Session Due to
Charging IPG Timeout
If the duration of the charging session exceeds 5 hours ±
5 minutes, your Vesta Charger will emit 3 long beeping
tones and display the Charging IPG Timeout Error screen,
indicated by a flashing hourglass icon in the center of the
screen (see Figure 20). Your Charger will then shut off. If
this should occur, please wait for approximately 10
minutes before initiating a new charging session.
Figure 20: Charging IPG Timeout Error Screen
6.8.4 Termination of Charging Session Due to Low
Charger Battery Level
If the battery charge level of your Vesta Charger drops
below 10% during a charging session, your Vesta
Charger will emit 3 long beeping tones and display the
Low Charger Battery Alert screen, indicated by an empty
battery icon with a flashing “X” over it (see Figure 21).
Your Charger will then shut off. If this occurs, recharge
the battery of your Vesta Charger as described in Section
6.6.
Figure 21: Low Charger Battery Alert Screen