35
power supply
input lines as
defined in
IEC 61000-4-
11
25/30 cycles
Interruptions:
100% reduction for
250/300 cycles
25/30 cycles
Interruptions:
100% reduction for
250/300 cycles
Charger requires
uninterrupted operation
during power mains
interruptions, it is
recommended to power the
Vesta Charger from an
uninterruptible power supply.
Power line
frequency
magnetic
fields (50/60
Hz) as defined
in IEC 61000-
3 A/m 3 A/m Power line frequency
magnetic fields (50/60 Hz)
should be at levels expected
in a typical home healthcare,
business, or hospital
environment.
Conducted RF
as defined in
IEC 61000-4-
6:2013
3 V r.m.s outside
industrial,
scientific, and
medical (ISM) and
amateur radio
bands between
0.15 MHz and
80 MHz, 6 V r.m.s.
in ISM and
amateur radio
bands between
0.15 MHz and
3 V r.m.s outside
industrial,
scientific, and
medical (ISM) and
amateur radio
bands between
0.15 MHz and
80 MHz, 6 V r.m.s.
in ISM and
amateur radio
bands between
0.15 MHz and
Portable and mobile RF
communications equipment
should be used no closer to
any part of the device,
including cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance:
d = 1.17√P
d = 1.17√P 80 MHz to
800 MHz
d = 2.33√P 800 MHz to
2.5 GHz
Where “P” is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and “d” is the
recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey, “a” should be less
than the compliance level in
each frequency range “b”.
Interference may occur in the
vicinity of equipment marked
with the following symbol:
Radiated RF
as defined in
IEC 61000-4-
3: 2006 +A1:
2007 +A2:
2010
10 V/m: 80 MHz to
2.7 GHz and
wireless
frequencies
10 V/m: 80 MHz to
2.7 GHz and
wireless
frequencies