48
ETSI EN 301 839
GUIDELINES AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC
EMMISSIONS OF THE OPTIMIZER SMART MINI IPG PURSUANT TO:
ETSI EN 301 839 V2.1.1 - Ultra Low Power Active Medical Implants (ULP-AMI) and
associated Peripherals (ULP-AMI-P) operating in the frequency range 402 MHz to
405 MHz; Harmonised Standard covering the essential requirements of article 3.2
of the Directive 2014/53/EU
The OPTIMIZER Smart Mini Implantable Pulse Generator, part of the OPTIMIZER Smart
Mini System is intended for use in an electromagnetic environment as specified below.
The patient implanted with the OPTIMIZER Smart Mini Implantable Pulse Generator must
ensure that it is used within the specified environment.
Emissions Test Compliance Electromagnetic
environment -
guidelines
Frequency Error
Complie
s with clause 5.3.1
The OPTIMIZER Smart
Mini IPG must emit
electromagnetic energy in
order to perform its
intended function when
communicating with the
Intelio Programmer or the
Vesta Charger. Nearby
electronic equipment may
be affected.
Occupied Bandwidt
h
Complies with clause 5.3.2
Power Output
Complies with clause 5.3.3
Transmitter Spurious
Emissions (30 MHz to
6
GHz)
Complies with clause 5.3.4
Frequency Stability Under
Low Voltage Conditions
Complies with clause 5.3.5
Spurious Radiation of
R
eceivers
Complies with clause 5.3.6