EasyManua.ls Logo

Impulse Dynamics OPTIMIZER Smart Mini - Page 56

Impulse Dynamics OPTIMIZER Smart Mini
63 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
50
Vesta Charger Wireless Nominal Specifications
Characteristic
Nominal
MICS MedRadio
Frequency Band 402 – 405 MHz Medical Implant
Communication Service (MICS)
Medical Device Radio Communication
Service (MedRadio)
Bandwidth
< 145 kHz
Modulation
FSK
Radiated Power
< 25 µW E.I.R.P.
Range
0 to at least 1.5 m
Transcutaneous Energy Transfer
Frequency Band 13.56 MHz
Industrial, Scientific, and Medical radio
band (ISM)
Bandwidth
< 0.014 MHz
Modulation Amplitude (slow to optimize coupling, no
data transmitted)
Radiated Power
< 0.6 W
Range
5
mm to 40
mm
OPTIMIZER Smart Mini IPG Wireless Nominal Specifications
Characteristic
Nominal
OPTI
link
MICS MedRadio
Frequency Band 402 – 405 MHz Medical Implant
Communication Service (MICS)
Medical Device Radio Communication
Service (MedRadio)
Bandwidth
< 145 kHz
Modulation
FSK
Radiated Power
< 25 µW E.I.
R.P.
Range
0 to at least 1.5 m

Table of Contents