FOR BEST ACCURACY
All tests that begin with 570 are for use with the eXact iDip
®
570 Smart
Photometer only.
Visit sensafe.com/idip/ for individual test instructions.
The meter has a default 3 minute automatic shutoff timer. You can change
the duration in the ‘Settings’ menu located within the eXact iDip
®
app.
Always ll the cell to capacity (4mL).
Test immediately after lling the cell with the water sample.
When testing pH, it is recommended to run the pH test prior to running
Chlorine. If you choose to run the pH after Chlorine, ensure you
thoroughly clean the cell with water and the brush provided.
To obtain optimal accuracy when testing outdoors (sunlight), use the
Cell Cover when zeroing and reading the sample.
Due to the strip slitting process, you may nd one or two strips that are
noticeably smaller or larger in width than the normal strips in the bottle.
These should be discarded. Using these strips may give unreliable results.
Meter is not compatible for use with powder pillows, tablets, and liquids
from other manufacturers.
Dip strip for entire countdown.
Each eXact
®
Strip Micro is valid for ONLY one test. Discard strip after use.
Dry the outside of the meter before storage.
Remove batteries before storing for prolonged periods.
Store the meter and test materials out of direct sunlight and away from
chemical storage areas.
Minimize exposure of meter and test reagents to heat above 90°F (32°C).
When installing batteries, make sure that the O-ring is still attached to
the screw before tightening.
To pair the eXact iDip
®
Smart Photometer with your smart device, open
the app and connect through the app. You cannot connect the devices
in the settings section of your smart device.
All results must be saved from the results page to history in order to
permanently save to your smart device.
The eXact iDip
®
Smart Photometer may experience Bluetooth
®
disconnection when submerged in water too deep. To avoid this,
collect water sample in a small container/cup and pour or pipette
(contact us for ordering) the water into the Sample Cell.
To ensure lab quality results, it is recommended to clean the cell with
the brush provided after every test.
Cleaning the cell with water and brush after each test and prior to
storage of unit is recommended. Use Distilled White Vinegar to clean
the cell after testing for Phosphate or Iron to remove reagent deposits
from the cell wall.