Guidance and manufacturer’s declaration – electromagnetic emissions
The System is intended for use in the electromagnetic environment specified below. The customer or the
user of the System should assure that it is used in such an environment.
Electromagnetic environment – guidance
The device uses RF energy only for its internal
function. Therefore, its RF emissions are very
low and are not likely to cause any interference
in nearby electronic equipment.
The device is suitable for use in all
establishments other than domestic and those
directly connected to the public low-voltage
power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/ flicker
emissions IEC 61000-3-3
Guidance and manufacturer’s declaration – electromagnetic immunity
The system is intended for use in the electromagnetic environment specified below. The customer or the
user should assure that it is used in such an environment.
Electromagnetic
environment – guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least 30
%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power supply
lines ± 1 kV for
input/output lines
± 2 kV for power
supply lines
Mains power quality should be
that of a typical commercial or
hospital environment.
± 1 kV differential mode
± 2 kV common mode
± 1 kV differential
mode ± 2 kV
common mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-
11
<5 % U
T
(>95 % dip in
U
T
) for 0,5 cycle
40 % U
T
(60 % dip in
U
T
) for 5 cycles
70 % U
T
(30 % dip in
U
T
) for 25 cycles
<5 % U
T
(>95 % dip in
U
T
) for 5sec
>95 % dip for 10 ms
60 % dip for 100 ms
30 % dip for 500 ms
95 % dip for 5000
ms
Mains power quality should be
that of a typical commercial or
hospital environment. If the
user of the device requires
continued operation during
power mains interruptions, it is
recommended that the device
be powered from an
uninterruptible power supply.
Power frequency
(50/60 Hz) magnetic
field IEC 61000-4-8
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE - U
T
is the AC mains voltage prior to application of the test level.