EasyManua.ls Logo

InMode VOTIVA - Recommended Separation Distances

InMode VOTIVA
51 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Votiva Operator Manual DO607094B 50
Recommended separation distances between
portable and mobile RF communications equipment and the System
The System intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the System can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the System as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum output
power of transmitter [W]
Separation distance according to frequency of transmitter [m]
150 kHz to 80 MHz

80 MHz to 800
MHz

800 MHz to 2,5 GHz
0,01
0.117
0.117
0.233
0,1
0.369
0.369
0.738
1
1.167
1.167
2.333
10
3.689
3.689
7.379
100
11.667
11.667
23.333
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.