16
Instrument Working Mode
INNO 320 has 4 working modes:
Wait mode
In wait mode, the instrument has been scheduled to record temperature. It is
waiting to enter to record mode and start temperature recording. The
symbol is displaying on the screen.
Users can set 3 methods to start recording:
1. Key start: Press and hold the key for 3 seconds, instrument will enter
record mode.
2. Time start: Set the time to start temperature recording. When the pre-
set time is reached, the instrument will enter record mode.
3. Temperature start: With temperature-rise start method, when the
measured temperature is higher than or equal to the pre-set value, then
the instrument will enter record mode. Similarly, with temperature-drop
start method, when the measured temperature is lower than or equal to
the pre-set value, then the instrument will enter record mode. In addition,
temperature-rise start and temperature-drop start method can be
enabled simultaneously.
Record mode
In record mode, instrument will record the measured temperature,
symbol is displaying in this mode.
The recording interval can be set between 10 seconds to 24 hours.
The instrument can stored up to 130,599 temperature values.
The upper and lower limit alarm functions can be set. When the recorded
temperature reaches the upper limit or reaches the lower limit, the
corresponding or symbol is displayed. If both the upper and lower
limits have been reached, both the and symbols will be displayed