17
simultaneously. If the LED alarm function is enabled, the red LED will flash
every 10 seconds after alarm was triggered.
Three recording method can be set:
1. Set a number of temperature values to be recorded, when the recording
amount was reached,recording will stop, instrument will enter end
mode.
2. Set a time period for recording,the recording start time and end time
should be set. When end time was reached, recording will stop,
instrument will enter end mode.
3. Cycle recording. Instrument will keeping recording data, when the
memory was full, the latest measured value will be recorded, and the
oldest value will be discarded. User can stop the recording by the
application software.
Note: If cycle recording method was used, the maximum and minimum
temperature values displayed on the instrument screen refer to the
highest and lowest values recorded by the instrument, not just the
highest and lowest values among the 130,599 temperature values
stored in the instrument. Therefore, the maximum and minimum values
displayed may no longer exist in the instrument's record.
End mode
In this mode, symbol is showing on the screen. Instrument was
completed the scheduled recording task, recording was stopped, but it will
still measure temperature and display it on the screen.
Measurement mode
In this mode, instrument will no record data, it is used as an accurate
temperature measuring instrument instead.
In this mode, when displaying maximum temperature value, the Max value
can be reset by pressing and hold the key for more than 3 seconds.
In this mode, when displaying minimum temp0erature value, the Min value
can be reset by pressing and hold the key for more than 3 seconds.