What causes excessive baseline movement in Innomed Medical Equipment?
- JJames AtkinsonSep 9, 2025
Excessive baseline movement in Innomed Medical Equipment can be caused by the amplifier’s too great time constant and high-impedance parts; it is possible, that the multiplexers before the ADC or the DC isolating capacitors are damaged. Check and repair the amplifier card according to the documentation. If necessary, replace the HS80G card. It may also be caused by condensing humidity. Check if it was caused by condensing humidity.