Why is there a memory error when I turn on my Innomed Medical Equipment device?
- CCheryl WattsSep 1, 2025
If your Innomed Medical Equipment device always starts with a memory error, the soldering may not be satisfactory. Check the soldering on the digital part of the Main Board and resolder the IC’s with the exception of the processor. If it cannot be repaired, replace the card. Also, the battery, or its connection / cable may be damaged. Check the battery. If it cannot be repaired, replace it. Check battery cabling.