EasyManua.ls Logo

Inogen One G3HF - Page 45

Inogen One G3HF
134 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
96-06434-00-01 revA
Service Manual, Inogen One G3HF
-45-
6. Seat the feet of the front housing into the corresponding grooves on the chassis (Figure
9) then push down and toward the device to secure the housing in the chassis.
Figure 9: Housing feet and corresponding chassis groove.
7. Place the device on its front side and repeat step 6 to secure the back housing into
the chassis.
8. Replace the two screws that connect the back housing to the chassis and the two screws
that connect the front housing to the chassis. Tighten screws to 3 in-lbs.
9. Return the concentrator to the upright position
10. Slide the battery back into the chassis. Ensure the battery clip is fully seated.