96-06434-00-01 revA
Service Manual, Inogen One G3HF
-53-
Figure 11: Housing feet and corresponding chassis grooves
2. Replace the two screws that connect the front housing to the chassis (Figure 2). Tighten
screws to 3 in-lbs.
3. Return the concentrator to the upright position.
4. Slide the battery back into the chassis. Ensure the battery clip is fully seated.