96-09302-00-01 revA DCR 19-030 21-Mar-2019 INOGEN ONE G5 TECHNICAL MANUAL
Page 10 of
18
7. Remove the dust caps from the new column assembly. Make sure there is no
dust or debris where the dust caps were located.
8. Insert the new column assembly into the concentrator.
9. Push the column assembly into the device until the spring loaded latch button
fully returns to the closed position.
10. Plug in external power. Do not power on the concentrator.
11. Press and hold the plus (+) and minus (-) buttons for 5 seconds. The screen
will display “sieve reset”.
12. Press the Mode button (bell symbol) once and screen will display “sieve reset
successful”.
13. Press the power button to turn on the concentrator and use normally.
3.2.4. Surface Cleaning
The outside case should be cleaned using a cloth dampened with a solution of mild liquid
detergent (such as Dawn
TM
) and water. Avoid getting water in or around the battery
connectors, power jack, or vents.
3.2.5. Battery Care and Maintenance
The Inogen One G5 Lithium Ion Battery requires special care to ensure proper performance
and long life. Use only Inogen One G5 Batteries with the Inogen One G5 Oxygen
Concentrator.
DESIGN
NOTE
The Inogen One G5 Oxygen Concentrator adjusts its oxygen production rate
to match the oxygen demand specified by the user flow setting. When the
device is used at lower settings, its battery life is extended. Additionally, at
lower flow settings, the concentrator does not generate as much heat and
noise, draw as much electric current from external power supplies, and many
system components do not wear as quickly.
3.2.5.1. Initial Battery State
The Inogen One G5 batteries are shipped partially charged. The battery can be used
immediately; however, we recommend the battery be charged upon receipt to
ensure adequate run time.
3.2.5.2. Normal Charging
CAUTION
Do not leave the column ends exposed after the dust caps are removed; they
should be inserted into the concentrator as soon as the dust caps have been
removed to minimize environmental exposure and maximize lifespan.
NOTE
Column change instructions are only to be used when maintenance is required.
Column should only be removed during this maintenance procedure.

Other manuals for Inogen One G5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Inogen One G5 and is the answer not in the manual?

Inogen One G5 Specifications

General IconGeneral
Battery Life (single battery)Up to 6.5 hours
Battery Life (double battery)Up to 13 hours
Operating Temperature41°F to 104°F (5°C to 40°C)
Storage Temperature-4°F to 140°F (-20°C to 60°C)
Altitude RangeUp to 10, 000 feet (3, 048 meters)
Flow Settings1-6
Weight (with single battery)4.7 pounds (2.1 kg)
PowerAC: 100-240V, 50-60Hz
Humidity Range0% to 95% non-condensing
Oxygen FlowPulse dose
Noise Level38 dBA