2.2. Click the Device dropdown box and select the device, select the COM Port (See Device Manager if unknown)
and open the correct amount of input channels (First = 1 and Last = total # of channels for device). For Model
9330, you will not be allowed to change the number of channels.
2.3. Click Connect
3. Each channel must now be scaled using the “SCALING” dialog box. Each channel must be scaled independently. If
the BSC8 was purchased with Interface load cells and a System Setup and Scaling then the scaling values will be
taken from the “Load Cell / BSC8 Digital Bridge Amplifier Calibration Certificate”
3.1. Physical full scale is typically the capacity of the sensor.
3.2. Electrical full scale output is the output of the sensor at the Physical full scale.
3.3. Input Range is always 2 mV/V and should not be changed.