EasyManuals Logo

Interface BSC4 Quick Start Guide

Interface BSC4
12 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #8 background imageLoading...
Page #8 background image
BSC4 Quickstart Guide
Interface Inc. 7401 East Butherus Drive, Scottsdale, Arizona 85260 USA Phone 480.948.5555 Fax 480.948.1924
www.interfaceforce.com Email: contact@interfaceforce.com 800.947.5598
Document Number 15-314 Rev A Page 8 of 12
2.2. Click the Device dropdown box and select the device, select the COM Port (See Device Manager if unknown)
and open the correct amount of input channels (First = 1 and Last = total # of channels for device). For Model
9330, you will not be allowed to change the number of channels.
2.3. Click Connect
3. Each channel must now be scaled using the “SCALING” dialog box. Each channel must be scaled independently. If
the BSC8 was purchased with Interface load cells and a System Setup and Scaling then the scaling values will be
taken from the “Load Cell / BSC8 Digital Bridge Amplifier Calibration Certificate”
3.1. Physical full scale is typically the capacity of the sensor.
3.2. Electrical full scale output is the output of the sensor at the Physical full scale.
3.3. Input Range is always 2 mV/V and should not be changed.

Other manuals for Interface BSC4

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Interface BSC4 and is the answer not in the manual?

Interface BSC4 Specifications

General IconGeneral
BrandInterface
ModelBSC4
CategoryAmplifier
LanguageEnglish