49
Integrated PASS
The S9 SCBA with Integrated PASS is the standard S9 default conguration. The S9
SCBA is NFPA 1981, 2019 and NFPA1982, 2018 certied including the integrated
PASS. As an option, the PASS can be turned OFF in the InMode program. When the
PASS is turned OFF, the S9 is certied to NFPA 1981, 2019 as an SCBA.
Start-up Self-Test
An automatic Start-up Self-Test option is available (start-up test default setting is OFF).
The automatic start-up test option can be turned ON in the InMode program by a
certied service technician. There are two start-up tests, a short test and a full test, that
are performed when the breathing apparatus is pressurized. The Short Test requires
no user interaction. The results of the test are presented on the Digital Display Unit
and must be acknowledged by the user before proceeding. The short test checks
cylinder pressure and the electronics and warning functions.
During the full test, motion sensor, panic button and the pressure tightness of the
breathing apparatus are checked with a minimum of user interaction. Note: The
Leakage and function test of the S9 (see section 4) and the checks during donning
(see section 5) still need to be performed by the User, even if the automatic Start-
up Self-Test option is activated. Generally the Start-up Self-Test option is used only
by a certied service technician, but can be activated for all Fire Department user’s
depending on the preference of the Fire Department.
11.2 INTERNAL PASS DATA LOGGING
The internal PASS data logging is accessible to Interspiro certied service technicians
in the InMode program (included as a minimum requirement of NFPA 1982, 2018
PASS data logging is not optional). The log records dates and times including:
- Power On and Power Off
· Pre-Alert and Lack of Motion Alarm
· Manual Alarm
· Reset
· Low Battery Warnings
In addition to PASS data, low air warnings are also included in the data log.