Troubleshooting
8Troubleshooting
8.1Identifyingandrepairingfaults
Thefollowingtableprovidesinformationonmalfunctionsandtheirpossiblecauses.Ifyoucannotcorrectthemalfunction
usingtheremediesnamed,pleasecontactyourdealerdirectly.
FaultPossiblecauseRemedy
Jacksocketofthehand
controlisnotlockedcorrectly.
Lockthejacksocket,→3.3Settingupthebathlift,page10.
Nomotornoiseaudible/lift
doesnotmove.
Batteryisat.Rechargethebatterycompletely,→3.4Chargingthebattery,
page13
.
Contactsaredirty.Cleanthecontacts,→contactyourdealer,ifnecessary.
Cableiscrushed/damaged.
Replacethehandcontrol,→contactyourdealer,ifnecessary.
Theliftcanberaisedbutnot
lowered.
Low-voltageprotectionis
activated.
Rechargethebatterycompletely,→3.4Chargingthebattery,
page13.
Theliftstopswhilebeing
raised.
Excessiveload.Observemaximumload,→9.1Dimensionsandweight,page
24.
Thebathliftisnotadhering
securelytothebathtub.
Suctioncupsareoldor
damaged.
Replacethesuctioncups,→contactyourdealer,ifnecessary.
Faultyelectronics.Haveyourauthorizeddealercheckthehandcontrol. Batteryisnotchargingand
theredindicatorlightashes
duringcharging
Faultybattery.Replacethehandcontrol.
Faultycharger.Haveyourauthorizeddealercheckthecharger.
Batteryistoohot.Donotchargethehandcontrolnearheatsources,e.g.heating
elements,indirectsunlight.Chargingcontinuesautomatically
oncethetemperaturedropstowithinpermissiblelimits.
1574470-D23