Aquatec®R/F/XL
10Electromagneticcompatibility(EMC)
10.1GeneralEMCinformation
MedicalElectricalEquipmentneedstobeinstalledandusedaccordingtotheEMCinformationinthismanual.
ThisproducthasbeentestedandfoundtocomplywithEMClimitsspeciedbyIEC/EN60601-1-2forClassBequipment.
PortableandmobileRFcommunicationsequipmentcanaffecttheoperationofthisproduct.
Otherdevicesmayexperienceinterferencefromeventhelowlevelsofelectromagneticemissionspermittedbytheabove
standard.Todetermineiftheemissionfromthisproductiscausingtheinterference,runandstoprunningthisproduct.If
theinterferencewiththeotherdeviceoperationstops,thenthisproductiscausingtheinterference.Insuchrarecases,
interferencemaybereducedorcorrectedbythefollowing:
•Reposition,relocate,orincreasetheseparationbetweenthedevices.
10.2Electromagneticemission
Guidanceandmanufacturer´sdeclaration
Thisproductisintendedforuseintheelectromagneticenvironmentspeciedbelow.Thecustomerortheuserofthisproduct
shouldassurethatitisusedinsuchanenvironment.
Emissionstest
ComplianceElectromagneticenvironment-guidance
RFemissions
CISPR11
GroupI
ThisproductusesRFenergyonlyforitsinternalfunction.Therefore,itsRF
emissionsareverylowandarenotlikelytocauseanyinterferenceinnearby
electronicequipment.
RFemissions
CISPR11
ClassB
Thisproductissuitableforuseinallestablishmentsincludingdomestic
establishmentsandthosedirectlyconnectedtothepubliclow-voltagepower
supplynetworkthatsuppliesbuildingsusedfordomesticpurposes.
261574470-D