Electromagneticcompatibility(EMC)
Harmonic
emissions
IEC61000-3-2
ClassA
Voltageuctuations
/ickeremissions
IEC61000-3-3
Complies
10.3ElectromagneticImmunity
Guidanceandmanufacturer´sdeclaration
Thisproductisintendedforuseintheelectromagneticenvironmentspeciedbelow.Thecustomerortheuserofthisproduct
shouldassurethatitisusedinsuchanenvironment.
Immunitytest
Test/Compliancelevel
Electromagneticenvironment–guidance
Electrostatic
discharge(ESD)
IEC61000-4-2
±8kVcontact
±2kV,±4kV,±8kV,
15kVair
Floorsshouldbewood,concreteorceramictile.Ifoorsarecoveredwith
syntheticmaterial,therelativehumidityshouldbeatleast30%.
Electrostatic
transient/burst
IEC61000-4-4
±2kVforpowersupply
lines;100kHzrepetition
frequency
±1kVforinput/output
lines;100kHzrepetition
frequency
Mainspowerqualityshouldbethatofatypicalcommercialorhospital
environment.
Surge
IEC61000-4-5
±1kVlinetoline
±2kVlinetoearth
Mainspowerqualityshouldbethatofatypicalcommercialorhospital
environment.
1574470-D
27