SECTION 1—GENERAL GUIDELINES
At’m 12 Part No. 1125035
Charging Batteries
DANGER
When using an extension cord, use only a three wire extension cord having at least
16 AWG (American Wire Gauge) wire and the same or higher electrical rating as
the device being connected. Use of improper extension cord could result in a risk of
fire and electric shock.
Three prong to two prong adapters should not be used. Use of three prong adapters
can result in improper grounding and present a shock hazard to the user.
NEVER attempt to recharge the batteries by attaching cables directly to the battery
terminals.
DO NOT attempt to recharge the batteries and operate the wheelchair at the same time.
DO NOT attempt to recharge the batteries when the wheelchair has been exposed to ANY
type of moisture.
DO NOT attempt to recharge the batteries when the wheelchair is outside.
DO NOT sit in the wheelchair while charging the batteries.
Read and carefully follow the manufacturer’s instructions for each charger (supplied or
purchased). If charging instructions are not supplied, consult a qualified technician for
proper procedures.
Ensure the pins of the extension cord plug are the same number, size, and shape as those
on the charger.
DO NOT under any circumstances cut or remove the round grounding plug from the
charger AC cable plug or the extension cord plug.
Grounding Instructions
DO NOT, under any circumstances, cut or remove the round grounding prong from any
plug used with or for Invacare products. Some devices are equipped with three-prong
(grounding) plugs for protection against possible shock hazards and fire. Where a
two-prong wall receptacle is encountered, it is the personal responsibility and obligation
of the customer to contact a qualified electrician and have the two-prong receptacle
replaced with a properly grounded three-prong wall receptacle in accordance with the
National Electrical Code. If you must use an extension cord, use only a three-wire
extension cord having the same or higher electrical rating as the device being connected.
In addition, Invacare has placed RED/ORANGE warning tags on some equipment. DO
NOT remove these tags.
Rain Test
Invacare has tested its power wheelchairs in accordance with ISO 7176 “Rain Test.” This
provides the end user or his/her assistant sufficient time to remove his/her power
wheelchair from a rain storm and retain wheelchair operation.