SECTION 5—WHEELCHAIR OPERATION
Part No. 1125035 27 At’m
Multi Function Charger Port
WARNING
Set-up/programming of the Electronic
Control Unit is to be performed only by
a qualified technician. The final adjust-
ments of the controller may affect other
activities of the wheelchair. Damage to
the equipment could occur under these
circumstances. If unqualified individuals
perform any work on these units, the
warranty is void.
FIGURE 5.2 Multi Function Charger Port
NOTE: For this procedure, refer to FIGURE 5.2.
The multi function charger port is located at the front of the joystick housing. This
provides easy access for charging the wheelchair batteries. This port also serves as the
Remote Programmer Communication connection.
Resetting the Circuit Breaker
WARNING
NEVER defeat or bypass the circuit breaker. Only replace with a circuit breaker of
the same rating.
NOTE: For this procedure, refer to FIGURE 5.3.
1. Remove the battery pack from the wheelchair base. Refer to Removing/Installing the
Battery Pack on page 36.
2. Open the battery pack by performing the following -
A. Grasp the side flaps and pull up.
B. Grasp the front flap and pull up.
3. To reset - press the circuit breaker reset button located inside the protector ring on top
of circuit breaker.
4. Close the battery pack by performing the following -
A. Fold down the front flap and press to engage the hook and loop straps.
B. Fold down the side flaps and press to engage the hook and loop straps.
NOTE: Ensure that the hook and loop straps are securely fastened before lifting the battery pack.
Multi-Function
Charger Port
Front View of Joystick