26
Number of
flashes
Fault Impact on
Scooter
Notes
4
Current Limit
Time Out
Drive inhibited
The scooter has drawn too much current
for too long, possibly because the motor
has been over-worked, jammed or
stalled. Turn the scooter power off, leave
it off for a few minutes, and then turn the
power back on again.
The controller has detected a short-
circuited motor. Check the cable harness
for short and check the motor: Contact
your authorised Invacare® Dealer.
5
Brake Fault Drive inhibited
Check that the declutching lever is in the
engaged position.
The park brake coil or wiring is faulty.
Check the park brake and wiring for open
or short circuits. Contact your authorised
Invacare® Dealer.
6
Out Of Neutral
At Power Up
Drive inhibited
Throttle is not in neutral position when
turning key switch on. Return throttle to
neutral, turn power off, and back on
again.
Throttle may need to be re-calibrated.
Contact your authorised Invacare®
Dealer.