ACCESSORIES
Intake Filter
Filter Cover
Owner's Manual
MS2100 Disposable Mobil-Mist
TM
Nebulizer (5 ml Capacity), w/Mouthpiece,
T-Piece, and 7 ft. tubing.
MS2200 Disposable Sidestream® Nebulizer (10 ml Capacity), w/Mouthpiece
and 7 ft. Tubing.
MS2400 Durable Sidestream® Nebulizer (10 ml Capacity), w/Mouth-
piece and 7 ft. Tubing. One (1) Year Limited Warranty.
IRC1198 TruZone® PFM Peak Flow Meter.
NOTE: See Invacare Dealer for additional Nebulizers.
NOTE: If using medical disinfectant cleaners, follow manufacturers instructions
carefully.
8. Keep the outer surface of the tubing dust-free by wiping regularly.
NOTE: The nebulizer air tubing does not have to be washed internally because
only filtered air passes through the tubing.
CLEANING OR REPLACING THE
FOAM INTAKE FILTER
NOTE: Clean the filter once a month or sooner if filter discolors.
1. Unplug unit.
2. Open cover of storage compartment to gain access to filter.
3. Remove filter and clean with soap and water. Let dry thoroughly
before use.
4. Replace filter if it becomes clogged, torn or has a worn appearance.
NOTE: Regular filter cleaning/replacing is necessary to help ensure proper
compressor performance.
WARNING
DO NOT run unit without filter.
MAINTENANCE/ACCESSORIES