12 TROUBLESHOOTING
Invacare
3G Storm Series® 126 Part No 1143151
Battery indicator flashes the charge level is
low - immediately after recharge.
Battery failure.
Malfunctioning battery charger.
Electrical malfunction.
Check batteries for shorted cell. Replace if necessary.
Contact Dealer/Invacare for Service.
Contact Dealer/Invacare for Service.
Battery indicator flashes the charge level is
low - too soon after being recharged.
Batteries not charged.
Weak batteries.
Have charger checked.
Replace batteries if necessary. Refer to Replacing Batteries
on page 94.
Motor “chatters” or runs irregular. Motor/gearbox malfunction. Stop use of Wheelchair. Contact Dealer/Invacare for Service.
Joystick erratic or does not respond as
desired.
Damaged motor coupling.
Electrical malfunction.
Controller programmed
improperly.
Contact Dealer/Invacare for Service.
Contact Dealer/Invacare for Service.
Contact Dealer/Invacare to have controller reprogrammed.
Wheelchair does not respond to commands. Electrical malfunction. Contact Dealer/Invacare for Service.
SYMPTOM PROBABLE CAUSE SOLUTIONS