1. Type of protection against electric shock: INTERNALLY POWERED EQUIPMENT.
2. Degree of protection against electric shock: TYPE BF APPLIED PART.
3. Mode of operation: CONTINUOUS OPERATION.
4. Equipment not suitable for category AP & APG equipment use in pressence.
STATEMENT
The system might not meet its performance specifications if stored or used outside the temperature and humidity as mentioned below.
Operating conditions: +5°C~+40°C 15%RH~93%RH 70kPa~106kPa
Storage conditions: -20°C~+55°C 0%RH~93%RH 50kPa~106kPa
troubleshooting
11
Error display Possible cause How to correct
Nothing is displayed when you push the
POWER button
No batteries are installed Insert the batteries.
The polarities of the batteries are
placed wrongly.
Insert the batteries correctly according to the
symbols.
Battery icon flashes
Batteries are low. Replace with new batteries.
E1: can't normally increase pressure
Check your cuff if there is any air
leakage.
E3: inflate pressure is too high Refasten the cuff and then measure again.