10
There are no user serviceable parts under the cover. Removal of the RF cover must
be done according to a procedure or damage could occur. (The RF cover edges are
RF sealed to the top heat sink, which makes it very difficult to slide off the cover.) The
laser may not operate properly without the cover in place. There are dangerous RF
voltages under the cover that will cause serious RF burns to skin if touched.
There are no user serviceable parts in the AC Power Supply.
The optical intensity at the output window is VERY high (>8500W/cm
2
). Any
contamination to the window surface could cause runaway destruction. To protect the
window, Iradion provides a power supply unit that includes a long life air pump and
filters to bathe the window in clean filtered air. Do NOT burn materials close to the
laser. The flow of air may not be enough to overcome the explosive vaporization of
objects near the output aperture.
The output aperture comes covered by a clearly marked dust
guard that must be removed before use. Firing the laser
through the cover WILL cause permanent destruction of the
laser.