15
Resistance
Range
100Hz
120Hz
1kHz 10kHz 100kHz
20.000Ω N/A 0.5% + 50
[1]
0.5% + 50
[1]
0.5% + 50
[1]
200.00Ω 0.5% + 8
[1]
0.5% + 8
[1]
0.5% + 8
[1]
0.5% + 8
[1]
2.0000kΩ 0.5% + 5 0.5% + 5 0.5% + 5 0.5% + 5
20.000kΩ 0.5% + 5 0.5% + 5 0.5% + 5 1.0% + 5
[2]
200.00kΩ 0.5% + 5 0.5% + 5 1.0% + 5
[2]
N/A
2.0000MΩ 0.5% + 5 1.0% + 5
[2]
N/A N/A
20.000MΩ 1.0% + 5
[2]
N/A N/A N/A
[1] Accuracy is specified after subtract of the offset resistance.
[2] < 50dgt rolling.
[3] If Q > 0.1, the accuracy should be multiplied by .
2
1 Q
DC Resistance
Range Resolution Accuracy
200.00Ω 10mΩ 0.5% + 8
[1]
2.0000kΩ 100mΩ 0.5% + 5
20.000kΩ 1Ω 0.5% + 5
200.00kΩ 10Ω 0.5% + 5
2.0000MΩ 100Ω 0.5% + 5
20.000MΩ 1kΩ 1.0% + 5
200.00MΩ 10kΩ 2.0% + 5
[2]
[1] Accuracy is specified after subtract of the offset resistance.
[2] < 50dgt rolling.