EasyManua.ls Logo

Isotech PLTC-782 - Taking Care of Your PLTC-782

Default Icon
43 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
V 1.16
30 | P a g e
7.2.1.1
Setting up the Reference Probe
To use the Automatic Cal/Adjust the calibration constants from your reference probe must be entered into the
PLTC-782 because the HP 3458A does not display in temperature values. The PLTC-782 will read the
resistance from the DMM and convert it to temperature internally via the calibration factors.
To enter the calibration data from your reference probe, enter the [PROBE SETUP] menu as shown below.
Select the type of coefficient data you have from your calibration certificate and enter the constants.
7.2.1.2
Set the DMM type and GPIB address
Enter the [GPIB SETUP] menu via the [SYSTEM MENU] as shown below. Select which DMM you have
connected and set the GPIB address.

Related product manuals