Remote Control
Copyright© Itech Electronic Co., Ltd. 10
administrator for server details. The same numbering notation applies as for the
IP Address. A value of 0.0.0.0 indicates that no default server is defined.
DNS is an internet service that translates domain names into IP addresses. It is
also needed for the instrument to find and display its hostname assigned by the
network. Normally, DHCP discovers the DNS address information; you only
need to change this if DHCP is unused or not functional.
DNS2: This field enters the secondary address of the server. Contact your LAN
administrator for server details. The same numbering notation applies as for the
IP Address. A value of 0.0.0.0 indicates that no default server is defined.
Socket Port: This value indicates the port number corresponding to the
Raw-socket service.
LAN Server
The configurable services include: mDNS, Ping, Telnet, Web, VXI and Raw
Socket.
The operation steps to configure are as follows.
This configures the instrument address (IP Mode).
1. Press [Shift]+[I-set] (Menu) to enter into the menu setting interface.
2. Use left/right key or rotate the knob to select Config > Communication
and press [Enter] to confirm.
3. Use left/right key or rotate the knob to select LAN and press [Enter] to
confirm.
4. Use left/right key or rotate the knob to select Config and press [Enter]
to confirm.
The first displayed menu item IP Mode is to configure the instrument
address.
5. Press [Enter] to confirm.
–
Auto: automatically configure the addressing of the instrument;
–
Manual: manually configure the addressing of the instrument.
6. After the setting is completed, press [Esc] to exit.
7. Restart the instrument and the modified configuration item will take
effect.
This selects the LAN services to enable or disable (Server Config).
1. Press [Shift]+[I-set] (Menu) to enter into the menu setting interface.
2. Use left/right key or rotate the knob to select Config > Communication
and press [Enter] to confirm.
3. Use left/right key or rotate the knob to select LAN and press [Enter] to
confirm.
4. Use left/right key or rotate the knob to select Config and press [Enter]
to confirm.
5. Use left/right key or rotate the knob to select Server Config and press
[Enter] to confirm.
6. Use left/right key or rotate the knob to select the desired service and
press [Enter] to confirm.
7. Rotate the knob to adjust the value and press [Enter] to confirm.
–
On: indicates enable the service
–
Off: indicates disable the service
8. After the setting is completed, press [Esc] to exit.
9. Restart the instrument and the modified configuration item will take
effect.
Reset the LAN to the Default Settings
The operation steps to reset the LAN to the default settings are as follows.
1. Press [Shift]+[I-set] (Menu) to enter into the menu setting interface.