DIGISPENSE 3009 CONTROLLER MODULE, STYLE B Page 3-107
29 October, 2019 IVEK Corp. P03R017E
Table 3.9 Common Operational Problems And Solutions - Cont.
PROBLEM
PROBABLE CAUSE
POSSIBLE SOLUTION
Unable to Erase At Least One
Backup Recipe
An error occurred when
resetting parameters.
Cycle power and attempt to reset parameters again.
Unable to Erase Any Recipes
An error occurred when
resetting parameters.
Cycle power and attempt to reset parameters again.
Backup Parameter Read Fault
The parameters backup
copy is corrupted
Clear fault. Check all parameter values. Cycle power.
Check all parameter values.
Parameter Value Out of Bounds
Fault
A value in the saved
parameters is out of
bounds of the system
Clear fault. Check all parameter values. Cycle power.
Check all parameter values.
Parameter Version Fault
The saved parameters
contain an unexpected
version number
Clear fault. Check all parameter values. Cycle power.
Check all parameter values.
Reset parameters (WARNING: will reset all recipes)
Parameters Blank Fault
The stored parameters
are corrupted.
Clear fault. Check all parameter values. Cycle power.
Check all parameter values.
Reset parameters (WARNING: will reset all recipes).
Cycle power.
Backup Configuration Read Fault
An error exists in backup
copy of the configuration
data
Reset parameters (WARNING: will reset all
parameters and recipes)
The stored configuration
data
is corrupted.
Cycle power.
Configuration Read Fault
The nonvolatile memory
is corrupted.
Reset parameters (WARNING: will reset all
parameters and recipes)
Configuration Value Out of Bounds
Fault
A stored configuration
data is corrupted
Cycle power.
Reset parameters (WARNING: will reset all
parameters and recipes)
Configuration Version Fault
The stored configuration
data
has an unexpected
version
Cycle power.
Reset parameters (WARNING: will reset all
parameters and recipes)
If none of the above solves the problem, contact
IVEK technical support for assistance.