OPERATING INCIDENTS
The suction cup does not adhere to the
component.
Deficient aspiration, Vacuum.
1 Verify if the suction cup is well placed and in
perfect condition.
JBC reserves the right to make technical changes without
prior notification.
Error messages
Whenever the ERR message appears, the
equipment turns itself off completely.
The following messages are displayed:
- BLANK SCREEN
Power failure.
Check for blown fuses.
- ERR 1
The temperature will not rise.
Possible causes: heating element open or blown
fuse.
Filter
Ref. 0861800
2 Check the incoming air filter in the interior of
the station and replace it if dirty or obstructed.
- ERR 3
No reading from the thermocouple.
Possible cause: open thermocouple.
- ERR 8
Faulty reading of the rotationsmeter of the airpump.
Possible causes: airpump damaged or faulty
function of the optical sensor circuit.
To recuperate any of these errors actuate the
general switch at the back of the station, the
pedal should not be pushed at this moment.