MANUAL DE INSTRUCCIONES CODIGO 80368 REV A DIC2018 (Sujetas a modicaciones sin previo aviso) Pag.: 15
J.P. SELECTA s.a.u. Autovía A2 Km 585.1 Abrera 08630 (Barcelona) España Tel 34 93 770 08 77 Fax 34 93 770 23 62
e-mail: selecta@jpselecta.es - website: www.jpselecta.es
GUARANTEE
This product is guaranteed for one year. The guarantee does not cover
damages caused by incorrect use or causes beyond the control of J.P.
SELECTA, s.a.u.
Any manipulation of the equipment by personnel not authorized by J.P.
SELECTA, s.a.u. automatically cancels the guarantee.
SPARE PARTS
To guarantee the safety of the equipment, all spare parts must be
purchased from J.P. SELECTA, s.a.u.
ADJUSTMENTS
The offset (A) and the ºC/ºF unit (B) can be adjusted.
A. Adjust the probe reading correction (offset)
Press keys (4) and (5) continuously and at the same time (about 5
seconds). The OFFs message will appear.
Use keys (4) / (5) to set the actual temperature reading value. Press
key (3) to conrm.
B. Presentation unit: ºC / ºF
Press key (6). The message Unit will appear. Using keys (4) / (5),
select the unit: ºC / ºF. Conrm with key (3).
Descripon Code
Model INCUBAT 2000260 2000261 2000262 2000263 2000264
Schuko connector 7171
Gasket 21254
Heang element 39122 39118 39120 39119 39121
Thermostat 38500
Glass door 47180 47181 47182 47183 47184
Safety thermostat 43153
Descripon Code
Model CONTERM 2000250 2000251 2000252 2000253 2000254
Schuko connector 7171
Gasket 21062
Heang element 39122 39118 39120 39119 39121
Thermostat 38500
Glass door -- -- -- -- --
Safety thermostat 43030