DOC. NO.: JDP-013-CUBEX
Issued Date: 2017.07.21
Revision: 2017.07.21(Rev.0)
19/54 Page
1.6.4 Intended conditions of use
a) Environment :
- General :
- Hospital use only
- Indoor use only
- Physical :
- System Storage/Transport Environmental Condition
Temperature range: -15°C ~ 55°C
Relative humidity range: 20% ~ 90% (non-condensing)
Atmospheric Pressure range: 50kPa ~ 106kPa
- System Operation Environmental Condition
Temperature range: 10°C ~ 35°C
Relative humidity range: 30% ~ 85% (non-condensing)
Atmospheric Pressure range: 70kPa ~ 106kPa
b) Frequency of use
- up to 100 times a day
c) Mobility
- Stationary only
d) Other conditions
- Not in the place exposed to toxic gas
- Not in the place humidity is high
- Not in the place exposed to steam
- Not in the place water drips
- Not in the place too much dust or sand exist
- Not in the place exposed to oil or fat
- Not in the place exposed to salty air
- Not in the place exposed to explosive gas or dust
- Not in the place excessive vibration or impact exists
- Not in the place the floor is inclined by 0.3 degree or more
- Not in the place power supply voltage changes abnormally
- Not in the place the load of power supply falls down or rises excessively
- Not in the place exposed to direct sunlight