19514822-V8.doc
English
Operating manual
Recirculating Coolers
FL1201
FL1203
FL1701 FLW1701
FL1703 FLW1703
1.951.4822-V8 11/17 Printed: 13.11.2017
w ww . . co m
information@itm.com1.800.561.8187

Questions and Answers

  • M
    Michael MurphyAug 16, 2025
    What to do if the Julabo Accessories recirculating cooler condenser cooling is affected?
    • B
      Brenda BrightAug 16, 2025
      If the cooling of the condenser in your Julabo Accessories is affected, you should clean the air-cooled condenser or check the flow rate and cooling water temperature on the water-cooled condenser.
  • B
    Brian MuellerAug 19, 2025
    What to do if the return temperature is above the high temperature protection switch-off value (85°C) in Julabo Accessories?
    • D
      David GomezAug 19, 2025
      If the return temperature exceeds the high temperature protection switch-off value (85°C) in your Julabo Accessories, examine the application's dimensions. If required, consider using a more powerful recirculating cooler.
  • A
    Ashley TranAug 22, 2025
    What to do if tube breakage occurred in Julabo Accessories?
    • D
      Derrick SextonAug 22, 2025
      If tube breakage occurs in your Julabo Accessories (leading to insufficient filling level due to excessive bath fluid being pumped out), replace the tubing and replenish the bath tank with bath fluid.
  • J
    Jacob FischerAug 25, 2025
    What to do if the alarm light illuminates and a continuous signal tone sounds on my Julabo FL1201 Accessories?
    • C
      Craig AllenAug 26, 2025
      If the alarm light is on and you hear a continuous signal, several factors could be the cause: * Press enter to quit the audible signal. * The recirculating cooler might be operating without enough bath fluid. Replenish the bath tank. * A tube breakage might have occurred, leading to insufficient filling. Replace the tubing and replenish the bath tank. * The return temperature may be too high. If necessary, use a stronger recirculating cooler. * The condenser cooling may be affected. Clean the air-cooled condenser. * Check the flow rate and cooling water temperature on water-cooled condenser. After fixing the issue, switch the mains power off and on again to cancel the alarm. If the problem persists, contact an authorized JULABO service station.
  • M
    Matthew SmithAug 27, 2025
    Why is the alarm light on and beeping on my Julabo FL1201?
    • L
      Lawrence SmithAug 27, 2025
      The alarm light and continuous beeping on your Julabo Accessories can indicate several issues: * Press enter to quit the audible signal. * The bath fluid level might be low. Refill the bath tank. * There could be a tube breakage. Replace the tube and refill the bath tank. * The return temperature might be too high, exceeding 85°C. Consider using a more powerful recirculating cooler. * The condenser might not be cooling properly. Try cleaning the air-cooled condenser. * Check the flow rate and cooling water temperature on water-cooled condenser. To reset the alarm, turn the power off and then on again after addressing the cause. If the problem continues, contact a JULABO service station.
  • S
    Stacey FranklinAug 30, 2025
    How to stop the alarm light and tone on Julabo FL1201 Accessories?
    • O
      Oscar RichardsAug 30, 2025
      To stop the alarm light and tone on your Julabo Accessories, consider these potential causes and solutions: * Press enter to quit the audible signal. * Check and replenish the bath fluid level in the recirculating cooler. * Inspect for tube breakage; replace tubing and replenish bath fluid if needed. * If the return temperature exceeds 85°C, consider using a stronger recirculating cooler. * Clean the air-cooled condenser if cooling is affected. * Check the flow rate and cooling water temperature on water-cooled condenser. After resolving the issue, switch the power off and on to clear the alarm. If the problem persists, contact a JULABO service station.
  • L
    Leslie ParkSep 2, 2025
    What does it mean when the alarm light comes on with a signal tone on Julabo FL1201?
    • K
      Kimberly EllisonSep 2, 2025
      The alarm light and signal tone on your Julabo Accessories indicates a problem. Possible causes include: * Press enter to quit the audible signal. * Insufficient bath fluid: Replenish the bath tank. * Tube breakage: Replace the tube and replenish the bath fluid. * High return temperature: Use a stronger recirculating cooler if necessary. * Affected condenser cooling: Clean the air-cooled condenser. * Check the flow rate and cooling water temperature on water-cooled condenser. After addressing the cause, reset the unit by turning the power off and on. If the alarm persists, contact a JULABO service station.
  • J
    John DixonSep 4, 2025
    What should I do if my Julabo FL1201 alarm light is on?
    • P
      philip42Sep 5, 2025
      If the alarm light on your Julabo Accessories is illuminated, along with a continuous signal tone, it could be due to: * Press enter to quit the audible signal. * Low bath fluid level. Replenish the bath tank. * A broken tube. Replace the tubing and replenish the bath tank. * Return temperature exceeding the limit. Use a more powerful recirculating cooler if needed. * Condenser cooling issues. Clean the air-cooled condenser. * Check the flow rate and cooling water temperature on water-cooled condenser. Turn the power off and on after correcting the problem to clear the alarm. If it continues, contact an authorized JULABO service station.
  • A
    Alexis PrinceSep 7, 2025
    What to do if the Julabo FL1201 Accessories compressor and circulating pump is shutdown?
    • S
      smithmitchellSep 7, 2025
      If the compressor and circulating pump of your Julabo Accessories are shut down and the alarm light is on, replenish the bath tank with the bath fluid.
  • L
    Lisa GarciaSep 10, 2025
    What to do if the return temperature of Julabo FL1201 is too high?
    • D
      Debra DiazSep 10, 2025
      If the return temperature of your Julabo Accessories is above the switch-off value, consider using a stronger recirculating cooler.