EasyManua.ls Logo

Julabo FL1201 - Autostart on; off; Remote Control: Activate - Deactivate

Julabo FL1201
34 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
24
7.7. AUTOSTART ON / OFF
The recirculating cooler has been configured and supplied by JULABO
according to N.A.M.U.R. recommendations. This means for the start
mode, that the unit must enter a safe operating state after a power
failure (non-automatic start mode). This safe operating state is
indicated by „OFF“ on the LED temperature display. A complete
shutdown of the main functional elements such as compressor and
circulating pump is effected simultaneously.
Should such a safety standard not be required, the AUTOSTART
function (automatic start mode) may be activated, thus allowing the
start of the circulator directly by pressing the mains power switch or
using a timer.
Keep depressed enter
and
turn on the unit with the mains power switch.
For a short while the LED DISPLAY indicates the effective start mode:
ÖAUTOSTART on. ÖAUTOSTART off.
Warning:
For supervised or unsupervised operation with the AUTOSTART function, avoid any
hazardous situation to persons or property.
The circulator does no longer conform to N.A.M.U.R. recommendations.
7.8. Remote control: activate – deactivate
(Interface OFF)
(Interface On)
Ð
Ð
Ð
Ð
The recirculating cooler is to be prepared for remote control by a
personal computer via the serial interface RS232.
Set the interface item from >IOFF< to >ION<.
Remote control: activate – deactivate:
x Switch off recirculating cooler by pressing the mains switch and
wait approx. 5 seconds.
x Keep depressed the keys
and enter simultaneously and
turn on the unit with the mains power switch.
>I OFF< No remote control via RS232 (Factory setting)
>I On< Remote control via RS232
L The software version and the type of unit is indicated
(see example on the left).
The display "rOFF" indicates the unit is ready to be operated via
remote control.
w ww . . co m
information@itm.com1.800.561.8187

Related product manuals